Frequency
10 MHz – 110 GHz

Lightwave component analyzers combining a VNA with optical transmitter and receiver modules for E/E, E/O, O/E and O/O measurements from 10 MHz up to 110 GHz.

Key highlights

  • Modulation frequency coverage from 10 MHz up to 110 GHz (6433P)
  • Four test modes: E/E, E/O, O/E and O/O
  • 1 Hz frequency resolution
  • Relative frequency response accuracy from ±0.7 dB (6433D)
  • Two- or four-port configurations for single-ended and balanced devices
  • Automatic fixture removal for on-wafer optical chip testing

Supplied by SeaGOAT Electronics, an independent international supplier of electronic test and measurement equipment. SeaGOAT Electronics does not manufacture this equipment; it is sourced through our manufacturing and supply partners.

Specifications

Frequency
Modulation frequency range6433D: 10 MHz – 20 GHz 6433F: 10 MHz – 43.5 GHz 6433H: 10 MHz – 50 GHz 6433L: 10 MHz – 67 GHz 6433P: 10 MHz – 110 GHz
Modulation frequency range (6433N)10 MHz – 90 GHz† Value from the SeaGOAT catalog. Manufacturer confirmation pending.
Frequency resolution1 Hz
Frequency accuracy (6433D/F/H/L)±1 × 10⁻⁷
Optical
Operating wavelength1310 nm ±20 nm, 1550 nm ±20 nm
External source input (6433D/F/H/L)1260 nm – 1610 nm
Average optical output power6433D: -2 dBm – +4 dBm 6433F/H/L: -2 dBm – +4 dBm (1310 nm), -1 dBm – +5 dBm (1550 nm) 6433P: -2 dBm – +3 dBm (1310 nm), -1 dBm – +5 dBm (1550 nm)
Average output power accuracy±0.5 dB
Maximum safe average input power6433D/F/H/L: input 1 +7 dBm, input 2 +17 dBm 6433P: +7 dBm
Average input power measurement range-35 dBm – +5 dBm
Average input power measurement accuracy±0.5 dB
Measurement
Test modesE/E, E/O, O/E, O/O
Relative frequency response accuracy6433D: ±0.7 dB 6433F: ±1.6 dB 6433H: ±1.6 dB 6433L: ±2.2 dB 6433P: ±2.7 dB
Absolute frequency response accuracy6433D: ±1.8 dB 6433F: ±2.0 dB 6433H: ±2.5 dB 6433L: ±2.7 dB
Frequency response repeatability6433D: ±0.3 dB 6433F: ±0.8 dB 6433H: ±1.0 dB 6433L: ±1.1 dB 6433P: ±1.4 dB
Group delay accuracy6433D: ±13 ps 6433F: ±18 ps 6433H: ±18 ps 6433L: ±31 ps
Phase accuracy6433D: ±3.5° 6433F: ±4.2° 6433H: ±4.2° 6433L: ±6.0°
Electrical ports2 or 4 (four-port option -400)
Display formatsLog/linear magnitude, phase, group delay, Smith chart, polar
Channels and windows (6433P)Up to 64 measurement channels, 32 windows, 16 traces per window
Software optionsAutomatic fixture removal (S07), time-domain measurement (S10), advanced time-domain analysis incl. TDR and eye diagram (S11), fast CW sweep (S18), frequency-offset measurement (S20)
General
Remote controlUSB, LAN; SCPI command set
Dimensions (W × H × D, without feet/handles)6433D/F/H/L: 426 mm × 400 mm × 600 mm 6433P main unit: 475 mm × 525 mm × 660 mm; extension module 97 mm × 73 mm × 236 mm; O/E module 97 mm × 81 mm × 215 mm
Maximum weight6433D/F/H/L: 63 kg 6433P: 80 kg
Maximum power consumption600 W

† Value from the SeaGOAT catalog. Manufacturer confirmation pending.

Overview

The 6433 series measures the modulation response of high-speed opto-electronic devices. It integrates four test modes in one instrument: electrical-to-electrical (E/E), electrical-to-optical (E/O), optical-to-electrical (O/E) and optical-to-optical (O/O). Models cover 10 MHz to 20 GHz (6433D), 43.5 GHz (6433F), 50 GHz (6433H), 67 GHz (6433L), 90 GHz (6433N) and 110 GHz (6433P).

Typical devices under test include electro-optic modulators, directly modulated lasers and transmitter optical assemblies (E/O), PIN and APD photodetectors and receiver assemblies (O/E), and passive optical parts such as fiber filters, attenuators and EDFAs (O/O). The analyzer reports bandwidth, magnitude and phase response and group delay, with log/linear magnitude, phase, group delay, Smith chart and polar display formats.

Guided electrical and optical calibration, one-button swept measurements and automatic fixture removal simplify on-wafer work with RF probes. Two-port and four-port configurations are available; the four-port versions measure differential gain and common-mode rejection of balanced transmitters and receivers. Operating wavelengths are 1310 nm and 1550 nm, and the 6433D/F/H/L accept an external source from 1260 nm to 1610 nm.

Applications

  • Bandwidth and S21/S11 testing of electro-optic modulators and directly modulated lasers
  • Frequency response and 3 dB bandwidth of photodetectors, ROSAs and TIA assemblies
  • Loss and flatness of passive optical components such as fiber filters
  • Differential gain and common-mode rejection of balanced optical transmitters and receivers
  • On-wafer testing of E/O and O/E chips with a probe station

Available models

Specifications differ between models. Select a model to include it in your inquiry.

ModelRangeModel-specific valuesContact Sales
6433D10 MHz – 20 GHz
Relative frequency response accuracy:
±0.7 dB
Absolute frequency response accuracy:
±1.8 dB
Frequency response repeatability:
±0.3 dB
Group delay accuracy:
±13 ps
Phase accuracy:
±3.5°
Average optical output power:
-2 dBm – +4 dBm
6433F10 MHz – 43.5 GHz
Relative frequency response accuracy:
±1.6 dB
Absolute frequency response accuracy:
±2.0 dB
Frequency response repeatability:
±0.8 dB
Group delay accuracy:
±18 ps
Phase accuracy:
±4.2°
Average optical output power:
-2 dBm – +4 dBm (1310 nm), -1 dBm – +5 dBm (1550 nm)
6433H10 MHz – 50 GHz
Relative frequency response accuracy:
±1.6 dB
Absolute frequency response accuracy:
±2.5 dB
Frequency response repeatability:
±1.0 dB
Group delay accuracy:
±18 ps
Phase accuracy:
±4.2°
Average optical output power:
-2 dBm – +4 dBm (1310 nm), -1 dBm – +5 dBm (1550 nm)
6433L10 MHz – 67 GHz
Relative frequency response accuracy:
±2.2 dB
Absolute frequency response accuracy:
±2.7 dB
Frequency response repeatability:
±1.1 dB
Group delay accuracy:
±31 ps
Phase accuracy:
±6.0°
Average optical output power:
-2 dBm – +4 dBm (1310 nm), -1 dBm – +5 dBm (1550 nm)
6433N10 MHz – 90 GHz—
6433P10 MHz – 110 GHz
Relative frequency response accuracy:
±2.7 dB
Frequency response repeatability:
±1.4 dB
Average optical output power:
-2 dBm – +3 dBm (1310 nm), -1 dBm – +5 dBm (1550 nm)
Required configuration:
6433-009 two-port frequency-extension controller, 6433-011 110 GHz coaxial extension modules and 6433P-S20 (four-port: 6433-010, four 6433-011 and 6433P-400)

Documents and datasheets

Original datasheets, manuals and application documents for this equipment are available from our sales team.

  • Brochure · ZH
Request the datasheet

Request a quote for the 6433D/F/H/L/N/P

Tell us your requirements and our sales team will reply with configuration options and a written quotation. The exact model number is included automatically. No prices or online ordering on this site.

Buying for a facility in the United States or Canada? How quotations, shipping and import documents work

We use your details only to answer this inquiry. No newsletters.

Specifications are compiled from the manufacturer’s official product information and the SeaGOAT catalog. Last verified November 17, 2026. Specifications may change; please confirm the exact configuration with our sales team before ordering. Partially verified against manufacturer data.

6317A tunable laser source, front view
Opto-Electronic Test InstrumentsBenchtop

6317A Tunable Laser Source

6317A

Mode-hop-free Littman external-cavity tunable laser for 1500 nm – 1600 nm (extendable to 1480 nm – 1680 nm), with ±5 pm absolute accuracy and sweeps up to 240 nm/s.

Wavelength
1480 – 1680 nm
Max. output
+13 dBm
View details
6362D optical spectrum analyzer, front view
Opto-Electronic Test InstrumentsBenchtop

6362 Series Optical Spectrum Analyzer

6362C/D/E/EA · 4 models

Benchtop optical spectrum analyzers with 20 pm minimum resolution bandwidth, covering 350 nm – 2500 nm across four models, for laser, WDM and optical amplifier testing.

Wavelength
350 – 2500 nm
View details
Request a Quote