Every VNA measurement is only as good as its calibration. A calibration kit contains standards with precisely known behavior, such as shorts, opens, loads, thrus or transmission lines, whose models the analyzer uses to solve for systematic errors in the test setup. After calibration, the reference plane moves to the end of the test cables or fixture, so the measured S-parameters describe the device rather than the cables and adapters in front of it.
Mechanical calibration kits are available in coaxial versions from DC to 110 GHz and in waveguide versions from 2.6 GHz to 500 GHz. The 2040X electronic calibration kits come as 2-port (20402) and 4-port (20404) modules in five frequency ranges, from 100 kHz–18 GHz up to 10 MHz–67 GHz. They connect by USB for both communication and power and have a maximum damage level of +10 dBm, so the RF ports must be protected from higher input levels.