Skip to content
SeaGOAT Electronics logo
4959 series handheld microwave analyzer, front three-quarter view (4959K shown)
Handheld RF & Microwave AnalyzersHandheld

4959 Series Handheld Microwave Analyzer

Model
4959B/D/E/G/H/K
Supplied by
SeaGOAT Electronics
Frequency
9 kHz – 54 GHz

Handheld microwave analyzer covering 9 kHz/300 kHz up to 54 GHz with spectrum analysis and dual-port vector network analysis, plus cable and antenna test options.

Key highlights

  • Six models up to 54 GHz: 9 kHz (spectrum) and 300 kHz (VNA) lower limits
  • Four-parameter dual-port VNA with 113 dB typical dynamic range (10 MHz – 20 GHz)
  • Spectrum analysis with -163 dBm/Hz typical DANL (2 MHz – 6 GHz, preamplifier on)
  • SSB phase noise of -110 dBc/Hz typical at 100 kHz offset, 1 GHz carrier
  • -20 °C to +55 °C operation, about 3 h typical battery life

Supplied by SeaGOAT Electronics, an independent international supplier of electronic test and measurement equipment. SeaGOAT Electronics does not manufacture this equipment; it is sourced through our manufacturing and supply partners.

Specifications

Spectrum analysis
Frequency range9 kHz – 54 GHz (by model)
Reference aging±0.5 ppm / year
Resolution bandwidth (RBW)1 Hz – 20 MHz (1-2-3-5-8 steps)
Video bandwidth (VBW)1 Hz – 20 MHz (1-2-3-5-8 steps)
SSB phase noise, 1 GHz carrier, 10 kHz offset≤ -108 dBc/Hz
SSB phase noise, 1 GHz carrier, 100 kHz offset≤ -110 dBc/Hz
SSB phase noise, 1 GHz carrier, 1 MHz offset≤ -118 dBc/Hz
SSB phase noise, 1 GHz carrier, 10 MHz offset≤ -129 dBc/Hz
DANL, 2 MHz – 6 GHz, preamplifier on (typ.)≤ -163 dBm/Hz
Second harmonic distortion, 50 MHz – 27 GHz≤ -70 dBc
Third-order intercept (TOI), 50 MHz – 54 GHz≥ +10 dBm
Total level uncertainty±1.30 dB (10 MHz – 20 GHz) ±1.80 dB (20 – 44 GHz) ±2.50 dB (44 – 50 GHz) ±2.80 dB (50 – 54 GHz)
Input attenuator0 – 30 dB, 2 dB steps
Real-time analysis bandwidth (option)40 MHz
Vector network analysis
Frequency range300 kHz – 54 GHz (by model)
Frequency accuracy±0.9 ppm
Frequency resolution1 Hz
IF bandwidth3 Hz – 100 kHz
Port output power, high power setting≥ -8.0 dBm (300 kHz – 10 MHz) ≥ +3.0 dBm (10 MHz – 9 GHz) ≥ 0.0 dBm (9 – 20 GHz) ≥ -2.0 dBm (20 – 26.5 GHz) ≥ -8.0 dBm (26.5 – 50 GHz) ≥ -12.0 dBm (50 – 54 GHz)
Port output power accuracy (-10 dBm output)±2.0 dB (300 kHz – 26.5 GHz) ±2.5 dB (26.5 – 50 GHz) ±3.0 dB (50 – 54 GHz)
Effective directivity≥ 42 dB (300 kHz – 4 GHz) ≥ 34 dB (4 – 20 GHz) ≥ 26 dB (20 – 54 GHz)
Effective load match≥ 30 dB (300 kHz – 4 GHz) ≥ 27 dB (4 – 26.5 GHz) ≥ 23 dB (26.5 – 54 GHz)
Reflection tracking±0.06 dB (300 kHz – 20 GHz) ±0.08 dB (20 – 50 GHz) ±0.12 dB (50 – 54 GHz)
Transmission tracking±0.08 dB (300 kHz – 4 GHz) ±0.16 dB (4 – 26.5 GHz) ±0.24 dB (26.5 – 50 GHz) ±0.28 dB (50 – 54 GHz)
System dynamic range≥ 90 dB (300 kHz – 2 MHz) ≥ 103 dB (2 MHz – 20 GHz) ≥ 93 dB (20 – 26.5 GHz) ≥ 90 dB (26.5 – 40 GHz) ≥ 81 dB (40 – 50 GHz) ≥ 80 dB (50 – 54 GHz)
Dynamic range, 10 MHz – 20 GHz (typ.)113 dB
Measurement functions
Standard functionsSpectrum analysis, vector network analysis
Optional functionsCable and antenna analysis (DTF, TDR), vector voltmeter, interference analysis (spectrogram, RSSI), channel scanner, field strength, USB CW and peak power measurement, USB OTDR and optical power meter, AM/FM/PM analysis, direction finding, 40 MHz real-time spectrum analysis, I/Q analysis, gated sweep
Interfaces
Other interfaces10 MHz reference in/out, external trigger in, GPS antenna (option), IF output (option), Wi-Fi antenna (option), all SMA female; LAN (RJ-45), USB 3.0 / USB-C, 3.5 mm audio, Micro SD
General
Display10.1-inch LCD with capacitive touch screen
Dimensions (W × H × D)316.7 × 236.5 × 75.0 mm (nominal, rear stand closed)
Power supplyAC adapter 100 – 240 V AC, 50/60 Hz (19 V DC, 4.7 A output); 10.8 V lithium-ion battery
Battery operating time3 h (typ.)
Operating temperature-20 °C – +55 °C
Storage temperature-50 °C – +70 °C

Overview

The 4959 series is a handheld microwave analyzer in six models with upper limits of 9, 18, 26.5, 44, 50 or 54 GHz. The spectrum analyzer starts at 9 kHz and the vector network analyzer at 300 kHz. Spectrum analysis and dual-port vector network analysis are standard; the instrument is battery powered, uses a 10.1-inch touch screen and operates from -20 °C to +55 °C.

The VNA measures all four S-parameters in magnitude and phase, supports time-domain analysis and mixed reflection S-parameters, and offers full two-port, unknown-thru, waveguide and external electronic calibration. Typical VNA dynamic range is 113 dB (10 MHz – 20 GHz). In spectrum mode, typical DANL is -163 dBm/Hz (2 MHz – 6 GHz, preamplifier on) and typical phase noise is -110 dBc/Hz at 100 kHz offset from 1 GHz.

Options add cable and antenna analysis with DTF and TDR, vector voltmeter, interference analysis, channel scanner, field strength, USB CW and peak power measurement, USB OTDR and optical power meter, AM/FM/PM analysis, direction finding, 40 MHz real-time spectrum analysis, I/Q analysis and gated sweep.

Applications

  • Field maintenance and support testing of electronic communication equipment
  • Antenna feeder SWR, return loss, insertion loss and impedance testing
  • S-parameter testing of amplifiers, filters, attenuators and duplexers
  • Transmitter and receiver subsystem testing in the field
  • Cable TV and cellular network spectrum and feed-line testing

Available models

Specifications differ between models. Select a model to include it in your inquiry.

ModelRangeModel-specific valuesContact Sales
4959B9 kHz – 9 GHz
DANL, preamplifier on (normalized to 1 Hz):
≤ -120 dBm (9 – 100 kHz) ≤ -140 dBm (100 kHz – 2 MHz) ≤ -159 dBm (2 MHz – 9 GHz) ≤ -156 dBm (9 – 14 GHz) ≤ -154 dBm (14 – 18 GHz)
Maximum safe input level (CW, ≥ 50 MHz):
+27 dBm
Test port:
Type N (female)
Weight:
≤ 3.7 kg (without battery)
4959D9 kHz – 18 GHz
DANL, preamplifier on (normalized to 1 Hz):
≤ -120 dBm (9 – 100 kHz) ≤ -140 dBm (100 kHz – 2 MHz) ≤ -159 dBm (2 MHz – 9 GHz) ≤ -156 dBm (9 – 14 GHz) ≤ -154 dBm (14 – 18 GHz)
Maximum safe input level (CW, ≥ 50 MHz):
+27 dBm
Test port:
Type N (female)
Weight:
≤ 3.7 kg (without battery)
4959E9 kHz – 26.5 GHz
DANL, preamplifier on (normalized to 1 Hz):
≤ -120 dBm (9 – 100 kHz) ≤ -140 dBm (100 kHz – 2 MHz) ≤ -159 dBm (2 MHz – 9 GHz) ≤ -156 dBm (9 – 14 GHz) ≤ -154 dBm (14 – 32 GHz) ≤ -152 dBm (32 – 40 GHz) ≤ -148 dBm (40 – 44 GHz) ≤ -145 dBm (44 – 50 GHz) ≤ -135 dBm (50 – 54 GHz)
Maximum safe input level (CW, ≥ 50 MHz):
+25 dBm
Test port:
3.5 mm (male)
Weight:
≤ 3.9 kg (without battery)
4959G9 kHz – 44 GHz
DANL, preamplifier on (normalized to 1 Hz):
≤ -120 dBm (9 – 100 kHz) ≤ -140 dBm (100 kHz – 2 MHz) ≤ -159 dBm (2 MHz – 9 GHz) ≤ -156 dBm (9 – 14 GHz) ≤ -154 dBm (14 – 32 GHz) ≤ -152 dBm (32 – 40 GHz) ≤ -148 dBm (40 – 44 GHz) ≤ -145 dBm (44 – 50 GHz) ≤ -135 dBm (50 – 54 GHz)
Maximum safe input level (CW, ≥ 50 MHz):
+25 dBm
Test port:
2.4 mm (male)
Weight:
≤ 3.9 kg (without battery)
4959H9 kHz – 50 GHz
DANL, preamplifier on (normalized to 1 Hz):
≤ -120 dBm (9 – 100 kHz) ≤ -140 dBm (100 kHz – 2 MHz) ≤ -159 dBm (2 MHz – 9 GHz) ≤ -156 dBm (9 – 14 GHz) ≤ -154 dBm (14 – 32 GHz) ≤ -152 dBm (32 – 40 GHz) ≤ -148 dBm (40 – 44 GHz) ≤ -145 dBm (44 – 50 GHz) ≤ -135 dBm (50 – 54 GHz)
Maximum safe input level (CW, ≥ 50 MHz):
+25 dBm
Test port:
2.4 mm (male)
Weight:
≤ 3.9 kg (without battery)
4959K9 kHz – 54 GHz
DANL, preamplifier on (normalized to 1 Hz):
≤ -120 dBm (9 – 100 kHz) ≤ -140 dBm (100 kHz – 2 MHz) ≤ -159 dBm (2 MHz – 9 GHz) ≤ -156 dBm (9 – 14 GHz) ≤ -154 dBm (14 – 32 GHz) ≤ -152 dBm (32 – 40 GHz) ≤ -148 dBm (40 – 44 GHz) ≤ -145 dBm (44 – 50 GHz) ≤ -135 dBm (50 – 54 GHz)
Maximum safe input level (CW, ≥ 50 MHz):
+25 dBm
Test port:
1.85 mm (male)
Weight:
≤ 3.9 kg (without battery)

Documents and datasheets

Datasheets for this product are available on request.

Request the datasheet

Request a quote for the 4959B/D/E/G/H/K

Tell us your requirements and our sales team will reply with configuration options and a written quotation. The exact model number is included automatically. No prices or online ordering on this site.

Buying for a facility in the United States or Canada? How quotations, shipping and import documents work

We use your details only to answer this inquiry. No newsletters.

Specifications are compiled from the manufacturer’s official product information and the SeaGOAT catalog. Last verified November 17, 2026. Specifications may change; please confirm the exact configuration with our sales team before ordering. Verified against manufacturer data.

4025 series handheld spectrum analyzer, front three-quarter view (4025D shown)
Handheld RF & Microwave AnalyzersHandheld

4025 Series Handheld Spectrum Analyzer

4025A/B/D/E/G/K · 6 models

New-generation handheld spectrum analyzer, 9 kHz up to 54 GHz in six models, with -165 dBm/Hz typical DANL, optional 120 MHz real-time analysis and 5G NR demodulation.

Frequency
9 kHz – 54 GHz
Real-time BW
120 MHz
View details
4957B handheld RF analyzer, front view
Handheld RF & Microwave AnalyzersHandheld

4957B Handheld RF Analyzer

4957B

Handheld 30 kHz – 6.5 GHz RF analyzer combining dual-port vector network analysis, cable and antenna test, spectrum analysis, vector voltmeter and power measurement.

Frequency
30 kHz – 6.5 GHz
View details
4957D/E/F handheld microwave analyzer, front view (4957E shown)
Handheld RF & Microwave AnalyzersHandheld

4957D/E/F Handheld Microwave Analyzer

4957D/E/F · 3 models

Handheld microwave analyzer up to 18, 26.5 or 40 GHz combining dual-port vector network analysis, cable and antenna test, spectrum analysis and power measurement.

Frequency
30 kHz – 40 GHz
View details
Request a Quote